A piece of health software can now fall under two product regulations at once. AI Act medical device compliance, under Regulation (EU) 2024/1689, does not replace MDR 2017/745: it sits on top of it. Many regulatory affairs teams discover the work while preparing a range extension or a certificate renewal. Here are the entry points, in the order in which we approach them.
First question: is your software a high-risk AI system?
Article 6(1) of the AI Act sets two cumulative conditions. The joint AIB 2025-1 / MDCG 2025-6 document, published in June 2025, restates them for medical devices:
- the AI system is itself a medical device, or it is a safety component of one within the meaning of Article 3(14);
- that device is subject to third-party conformity assessment by a notified body under MDR 2017/745 or IVDR 2017/746.
The consequence is immediate: the device class decides. The MDCG sets out a non-exhaustive reading:
- outside the high-risk category: class I non-sterile, with no measuring function and not a reusable surgical instrument; class A non-sterile IVDs; devices manufactured and used within the same health institution (Article 5(5) of the MDR);
- high risk: class I sterile, with a measuring function or a reusable surgical instrument; classes IIa, IIb, III; Annex XVI products; sterile class A IVDs and classes B, C and D.
Staying outside the high-risk category does not exempt you from the regulation's other obligations. Nor does the AI Act change the MDR or IVDR classification of the device (recital 51), and the prohibited practices in Article 5, applicable since 2 February 2025, and the transparency obligations in Article 50 apply whatever the class.
What the AI Act adds to the MDR file
The requirements for high-risk systems are set out in Articles 8 to 15. They partly overlap with the MDR, but they open angles that the technical documentation does not always cover explicitly:
- risk management (Article 9): an iterative process across the whole lifecycle, extended to fundamental rights;
- data and data governance (Article 10): training, validation and testing data sets that are representative, complete in view of the intended purpose and examined for possible bias;
- technical documentation (Article 11 and Annex IV);
- record-keeping (Article 12): automatic logging;
- transparency (Article 13): operation intelligible enough for the deployer to interpret outputs correctly;
- human oversight (Article 14): design measures that allow natural persons to oversee the system;
- accuracy, robustness and cybersecurity (Article 15).
The MDCG stresses a point of vocabulary with practical consequences: the "deployer" under the AI Act, the entity using the system under its authority, is not the "user" under the MDR, the healthcare professional or lay person handling the device. Instructions for use and training material must address both.
One technical documentation set, not two
Article 8(2) allows the manufacturer to integrate the testing, reporting, information and documentation required by the AI Act into what is already produced under the MDR or IVDR, and the MDCG explicitly encourages that route. Article 11(2) goes further: for a medical device, a single set of technical documentation must be drawn up, covering Annex II of the MDR and Annex IV of the AI Act.
The same logic applies to the quality management system: the AI Act requirements target the AI system specifically, they complement the existing QMS rather than duplicate it, and can be folded into it. In practice, this means deciding, procedure by procedure, where to place data governance, logging, human oversight and AI-specific post-market monitoring. It is also the moment to check that terminology stays stable from one document to the next, a control point common to all the regulatory frameworks we cover. Sampling rules, by contrast, do not move: the notified body's review of technical documentation follows the MDR and IVDR rules, including for classes IIa and IIb.
Conformity assessment remains the MDR route
For a high-risk system falling under Article 6(1), the applicable procedure is the MDR or IVDR one, under Article 43(3). There is no additional marking and no second notified body to appoint. The notified body designated under the MDR examines the requirements of Articles 8 to 15 and certain points of Annex VII of the AI Act, namely points 4.3, 4.4, 4.5 and the fifth paragraph of point 4.6, within the MDR procedure.
Watch the concept of substantial modification. It is autonomous, defined in Article 3(23), and triggers a fresh conformity assessment under Article 43(4). For systems that continue to learn after being placed on the market, changes pre-determined by the manufacturer, assessed at the initial conformity assessment and described in the technical documentation under Annex IV point 2(f), do not constitute a substantial modification. That change plan is written before the notified body review, not after.
The timeline after the digital omnibus
The AI Act entered into force on 1 August 2024. Its deadlines were revised by Regulation (EU) 2026/1744, published in the Official Journal of the European Union on 24 July 2026 and in force since 27 July 2026. The dates to keep:
- 2 February 2025: prohibited practices under Article 5;
- 2 August 2025: governance and obligations for general-purpose AI models;
- 2 August 2026: general application of the regulation;
- 2 December 2027: high-risk systems listed in Annex III;
- 2 August 2028: high-risk systems falling under Annex I, which includes medical devices.
That deferral (the medical device date was originally set at 2 August 2027) changes neither the requirements nor the logic of dual compliance. It only moves the point at which a notified body will hold you to them. MDR certification cycles run in months, so the margin is shorter than it looks.
Where to start with AI Act medical device compliance
Five steps, in this order.
- Qualify. Write down in plain terms whether your software is an AI system within the meaning of Article 3(1), and whether it meets the two conditions of Article 6(1). That note runs to two pages and underpins the rest.
- Map the gaps. Take Articles 8 to 15 one by one and identify, for each, the section of the MDR technical documentation that already answers it, and the one that is missing.
- Decide the documentation structure. A single set, yes, but with what numbering, what mapping table to Annex IV, what owner for each section?
- Open the subject with your notified body. Its expectations on Annex IV and on the pre-determined change plan are firming up.
- Update your internal procedures, down to the instructions for use and deployer training. Those in-house requirements can be submitted as a framework of your own.
In practice
Dual compliance is not a second file, it is an enriched MDR file reviewed against an additional grid. The difficulty is not legal, it is documentary: every requirement needs a written answer, vocabulary must stay stable from one document to the next, and no claim in the instructions for use should go beyond what the documentation demonstrates. That cross-review is what we support with EryonOne, where every finding quotes the exact passage and the reference text, with a confidence score, and where the decision to accept or dismiss stays with the expert. The report is working material, not an approval.
Bibliographie
- Regulation (EU) 2024/1689 of the European Parliament and of the Council of 13 June 2024 laying down harmonised rules on artificial intelligence, EUR-Lex, 2024
- Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices, EUR-Lex, 2017
- AIB 2025-1 / MDCG 2025-6, Interplay between the Medical Devices Regulation (MDR) & In vitro Diagnostic Medical Devices Regulation (IVDR) and the Artificial Intelligence Act (AIA), European Commission, June 2025
- Regulation (EU) 2026/1744 amending Regulation (EU) 2024/1689 (digital omnibus on artificial intelligence), EUR-Lex, 2026
- MDCG 2019-11, Guidance on Qualification and Classification of Software in Regulation (EU) 2017/745 and Regulation (EU) 2017/746, Medical Device Coordination Group, 2019